16.4 Field instrument: The flux-gate magnetometer
There are two classes of instruments to measure the ambient magnetic field: scalar and vector
magnetometers. While nuclear precession and optical pumping magnetometers are the most common scalar
magnetometers used on board s/c (see Pfaff et al.
, 1998b for related material), the flux-gate magnetometer
is, with no doubt, the mostly used one to perform vector measurements of the ambient magnetic field. In
this section, we will briefly describe only this last instrument just for those who are not familiar at all with
this kind of measurements in space.
The working principle of this magnetometer is based on the phenomenon of magnetic hysteresis. The
primary element (see Figure 106) is made of two bars of high magnetic permeability material. A
magnetizing coil is spooled around the two bars in an opposite sense so that the magnetic field created
along the two bars will have opposite polarities but the same intensity. A secondary coil wound around
both bars will detect an induced electric potential only in the presence of an external magnetic
field.
The field amplitude
produced by the magnetizing field
is such that the material periodically
saturates during its hysteresis cycle as shown in Figure 107.
In absence of an external magnetic field, the magnetic field
and
produced in the two bars will
be exactly the same but out of phase by
since the two coils are spooled in an opposite sense. As a
consequence, the resulting total magnetic field would be
as shown in Figure 107. In these conditions no
electric potential would be induced on the secondary coil because the magnetic flux
through the
secondary is zero.
On the contrary, in case of an ambient field
, its component parallel to the axis of the bar is
such to break the symmetry of the resulting
(see Figure 108).
represents an offset that would
add up to the magnetizing field
, so that the resulting field
would not saturate in
a symmetric way with respect to the zero line. Obviously, the other sensitive element would
experience a specular effect and the resulting field
would not be zero, as shown in
Figure 108.
In these conditions the resulting field
, fluctuating at frequency
, would induce an
electric potential
, where
is the magnetic flux of
through the secondary
coil.
At this point, the detector would measure this voltage which would result proportional to the
component of the ambient field
along the axis of the two bars. To have a complete measurement of
the vector magnetic field
it will be sufficient to mount three elements on board the spacecraft, like the
one shown in Figure 106, mutually orthogonal, in order to measure all the three Cartesian
components.